    Evaluation and Improvement of the Accuracy of TEM Transmission Cells with Internal Field Measurements

  • Author

    Jokela, K. ; Puranen, L.

  • Author_Institution
    Institute of Radiation Protection, P O Box 268, SF-00101 HELSINKI 10, FINLAND
  • fYear
    1983
  • fDate
    3-8 Sept. 1983
  • Firstpage
    652
  • Lastpage
    657
  • Abstract
    Two square TEM transmission cells with side heights 30 cm and 50 cm were designed for the calibration of RF-hazard monitors. The error due to reflections was reduced from ±10 % to ±2 % (input VSWR < 1.2) by lengthening straight sections and measuring the longitudinal E-field patterns at calibration frequencies. Comparison measurements with small probes showed that the maximum difference between the cells was 7 % except some sharp resonance points caused by TEM harmonics and TE01 mode. The vertical field non-uniformity was also examined.