• DocumentCode
    1990097
  • Title

    Retrieval Analysis of a Cementless Modular Total Hip Arthroplasty Prosthesis

  • Author

    Paliwal, Manish ; Allan, D. Gordon ; Filip, Peter

  • Author_Institution
    Coll. of New Jersey, Ewing
  • fYear
    2007
  • fDate
    14-17 Oct. 2007
  • Firstpage
    553
  • Lastpage
    558
  • Abstract
    This paper reports the analysis of a modular S-ROM hip implant, which was retrieved at revision, secondary to aseptic loosening approximately after 24 months of implantation. Light microscopy and scanning electron microscopy confirmed the presence of pitting, fretting corrosion, plastic deformation, and stress induced corrosion cracking. Energy dispersive X-ray microanalysis of the stem surface revealed the release of metal ions secondary to environmental interaction. High resolution inductively coupled plasma mass spectroscopy performed on the blood serum sample confirmed the elevated titanium ion levels (4.66 ppb). Finite element analysis of the implant showed that the micromotion and stress levels were the maximum at the proximal lateral region of the taper junction, in congruence with the observation. This case shows that taper performance is influenced by implant´s metallurgy and geometry. Patient weight, activity level, and effect of the local microenvironment also play a significant role.
  • Keywords
    X-ray chemical analysis; biomechanics; biomedical materials; blood; finite element analysis; mass spectroscopy; optical microscopy; plastic deformation; prosthetics; scanning electron microscopy; stress corrosion cracking; titanium alloys; wear; S-ROM hip implant; TiJkJk; activity level; blood serum; cementless modular total hip arthroplasty prosthesis; energy dispersive X-ray microanalysis; finite element analysis; fretting corrosion; inductively coupled plasma mass spectroscopy; light microscopy; local microenvironment effect; micromotion; patient weight; pitting; plastic deformation; retrieval analysis; scanning electron microscopy; stress induced corrosion cracking; Corrosion; Dispersion; Energy resolution; Hip; Implants; Plastics; Prosthetics; Scanning electron microscopy; Stress; Surface cracks; Retrieval analysis; arthroplasty; hip; microscopy; modular;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioinformatics and Bioengineering, 2007. BIBE 2007. Proceedings of the 7th IEEE International Conference on
  • Conference_Location
    Boston, MA
  • Print_ISBN
    978-1-4244-1509-0
  • Type

    conf

  • DOI
    10.1109/BIBE.2007.4375615
  • Filename
    4375615