Title :
Contrast-detail evaluation of a full-field digital mammography system
Author :
Darambara, D.G. ; Taibi, A. ; Speller, R.D. ; Gambaccini, M.
Author_Institution :
Dept. of Med. Phys. & Bioeng., Univ. Coll. London, UK
Abstract :
We have been investigating the potential of large-area flat-panel a-Si:H imaging devices for full-field digital mammography. The performance of a clinical mammographic system specially developed to provide geometric direct magnification equipped with an a-Si:H digital detector in conjunction with a series of Gd2O2S:Tb phosphor screens was evaluated in terms of image quality and entrance surface dose by using the contrast-detail CDMAM phantom over a range of exposure levels. We also compared the contrast-detail detectability properties of the a-Si:H-based digital imaging system with those of a conventional screen-film combination. The results from this study reveal the superiority of this prototype a-Si:H-based digital imaging system both in terms of contrast-detail resolution and significant entrance surface dose reduction in comparison with the conventional screen-film combination. These advantages are of particular significance in the breast imaging. Therefore, the combination of a-Si:H digital array with the direct magnification mammography system has the potential to fulfil the image quality requirements for full-field digital mammogarphy and to offer significant improvements in the efficient early detection of breast cancer
Keywords :
X-ray detection; amorphous semiconductors; diagnostic radiography; mammography; silicon radiation detectors; Gd2O2S:Tb phosphor screens; Si:H; breast cancer; breast imaging; clinical mammographic system; contrast-detail detectability properties; contrast-detail evaluation; contrast-detail resolution; direct magnification mammography system; entrance surface dose; full-field digital mammography system; geometric direct magnification; image quality; large-area flat-panel a-Si:H imaging devices; Breast; Cancer detection; Detectors; Digital images; Image quality; Image resolution; Imaging phantoms; Mammography; Phosphors; Prototypes;
Conference_Titel :
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location :
Seattle, WA
Print_ISBN :
0-7803-5696-9
DOI :
10.1109/NSSMIC.1999.842479