DocumentCode :
1997902
Title :
Digital alpha/beta pulse shape discrimination of CsI:Tl for on-line measurement of aqueous radioactivity
Author :
Tan, H. ; DeVol, T.A. ; Fjeld, R.A.
Author_Institution :
Dept. of Environ. Syst. Eng., Clemson Univ., SC, USA
Volume :
1
fYear :
1999
fDate :
1999
Firstpage :
499
Abstract :
A digital pulse shape discrimination data acquisition system developed around a commercially available digital oscilloscope card and data acquisition software was applied to on-line measurement of aqueous radioactivity. The radiation detector consisted of a flow-cell containing granular 63-90 μm Parylene C coated CsI:Tl and two Burle 8850 photomultiplier tubes (PMTs). The PMTs were connected in a software-realized coincidence mode. The digital data acquisition system consisted of a GaGe CompuScope 8012A/PCI 12-bit dual channel 50-MHz digital oscilloscope card residing in a Pentium III 450 MHz personal computer and the LabVIEW 5.1 data acquisition software. The individual fast anode scintillation pulses were digitized by the oscilloscope, transferred from on-board memory to computer memory, and processed by the LabVIEW Virtual Instrument. The coincidence resolving time was set to 100 ns to reduce background. Pulse shape discrimination was achieved using the charge integration technique. Experimental tests showed very good pulse shape discrimination between alpha and beta particles (FOM=1.65). The alpha and beta detection efficiencies for the CsI:Tl flow-cell were typically 31% and 26%, respectively. This system was also coupled to a liquid ion chromatography system for on-line measurement of aqueous alpha/beta radionuclides
Keywords :
alpha-particle detection; beta-ray detection; caesium compounds; computerised instrumentation; data acquisition; digital instrumentation; environmental science computing; pollution measurement; radioactive pollution; radioactivity measuring apparatus; solid scintillation detectors; thallium; Burle 8850 photomultiplier tubes; CsI:Tl; LabVIEW 5.1 data acquisition software; Parylene C coated CsI:Tl; aqueous radioactivity; charge integration technique; data acquisition software; digital alpha/beta pulse shape discrimination; digital oscilloscope card; digital pulse shape discrimination data acquisition system; fast anode scintillation pulses; flow-cell; liquid ion chromatography system; on-line measurement; software-realized coincidence mode; Anodes; Data acquisition; Microcomputers; Oscilloscopes; Photomultipliers; Pulse measurements; Pulse shaping methods; Radiation detectors; Shape measurement; Software measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location :
Seattle, WA
ISSN :
1082-3654
Print_ISBN :
0-7803-5696-9
Type :
conf
DOI :
10.1109/NSSMIC.1999.842535
Filename :
842535
Link To Document :
بازگشت