Title :
Research on detector of wide energy range neutron in BEPCII
Author :
Liu, Chunyu ; Li, Jianping ; Dong, Baozhong
Author_Institution :
Coll. of Nucl. Sci. & Technol., Harbin Eng. Univ., Harbin, China
Abstract :
In this paper according to the radiation environment of Beijing Electron Positron Collider (BEPC) synchrotron radiation hall, the neutron detector design is finished by using MCNP cord. Then the calibration and physics test of detectors are finished. Using the detector the measurements of radiation level in BEPC synchrotron radiation hall were carried out. The measurement of high energy dose rate is according with measurement result of method of ration of C/In. It provides good method to measure high energy neutron dose around high energy accelerator.
Keywords :
Monte Carlo methods; calibration; neutron detection; proportional counters; radiation monitoring; synchrotrons; BEPC synchrotron radiation hall; BEPCII; Beijing Electron Positron Collider; MCNP; detector calibration; detector physics testing; high energy accelerator; high energy dose rate; high energy neutron dose; radiation environment; wide energy range neutron detector; Calibration; Detectors; Energy measurement; Neutrons; Particle measurements; Radiation detectors; Sensitivity; accelerator; neutron detector; test; wide energy range;