• DocumentCode
    2002359
  • Title

    The relative impact of scatter on absolute myocardial perfusion quantitation: an EG84 Monte Carlo study

  • Author

    Tang, H. Roger ; Da Silva, Angela J. ; Hasegawa, Bruce H.

  • Author_Institution
    Dept. of Radiol., California Univ., San Francisco, CA, USA
  • Volume
    3
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    1224
  • Abstract
    Increased inferior activity in attenuation-corrected myocardial images has been attributed to scatter from the liver. To study this effect, the authors used EGS4 to simulate the SPECT acquisition of a 1.5-cm thick spherical shell placed inside a 24-cm diameter tank. The distance between the shell and a liver-simulating hemisphere was varied from 0.1 to 2 cm and the shell:liver activity concentration ratio was varied from 4:1 to 1:1 to estimate the impact of liver scatter. Profiles through the projection data demonstrated that about 20% of detected counts were scatter, with almost no bias on the side of the shell closest to the liver, except when the shell was very close to the liver. Resulting MLEM-reconstructed images showed only a very small bias proximal to the liver, with quantitative errors dominated by resolution effects. The authors´ modeled quantitation procedure underestimated the overall activity concentration by 15% but showed only small regional bias due to liver scatter
  • Keywords
    Monte Carlo methods; cardiology; gamma-ray scattering; haemorheology; image reconstruction; image resolution; liver; muscle; physiological models; 1.5 cm; 24 cm; EG84 Monte Carlo study; SPECT; absolute myocardial perfusion quantitation; liver scatter; liver-simulating hemisphere; medical diagnostic imaging; nuclear medicine; quantitative errors; resolution effects; spherical shell; Detectors; Electromagnetic scattering; Geometry; Heart; Imaging phantoms; Liver; Monte Carlo methods; Myocardium; Particle scattering; Single photon emission computed tomography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.842779
  • Filename
    842779