    Design and early testing of a new medium-energy transmission device for attenuation correction in SPECT and PET

  • Author

    Gagnon, D. ; Tung, C.H. ; Zeng, L. ; Hawkins, W.G.

  • Author_Institution
    Div. of Nucl. Med., Picker Int., Cleveland, OH, USA
  • Volume
    3
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    1349
  • Abstract
    A new concept of transmission device for SPECT and PET attenuation correction is described. The system uses a medium energy Barium source (main peak at 356 keV) in a fan beam geometry suitable for a variety of camera configurations. Advantages of Barium are numerous and seem to overcome some limitations observed in earlier generations of scanning devices. Furthermore, it is shown that with slight modifications of the acquisition electronics and imaging protocol, high quality anatomical information can also be obtained for registration purposes
  • Keywords
    biomedical equipment; gamma-ray absorption; positron emission tomography; single photon emission computed tomography; 356 keV; Ba; PET attenuation correction; SPECT; acquisition electronics; camera configurations; fan beam geometry; high quality anatomical information; imaging protocol; medical diagnostic imaging; medical instrumentation; nuclear medicine; scanning devices; Attenuation; Barium; Cameras; Collimators; Detectors; Geometry; Nuclear medicine; Personnel; Positron emission tomography; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.842806
  • Filename
    842806