DocumentCode
2003356
Title
Modeling the pulse-echo response of a 2D CMUT array element
Author
Yamaner, Feysel Yalcin ; Bozkurt, Ayhan
Author_Institution
Fac. of Eng. & Natural Sci., Sabanci Univ., Istanbul, Turkey
fYear
2009
fDate
20-23 Sept. 2009
Firstpage
1
Lastpage
4
Abstract
In this paper, we present an equivalent circuit model for a 2D CMUT array element to asses the overall pulse-echo response. The proposed circuit models the propagation medium, and incorporates reactive effects due to transducer dimensions which are comparable to the acoustic wavelength, and the effects of diffraction and loss. The model has been verified experimentally using hydrophone and pulse-echo measurements on a set-up that incorporates a high-voltage pulse generator and receive amplifier IC. The presented methodology can be used to evaluate the performance of a front-end IC within the EDA environment in which the circuit is designed.
Keywords
capacitive sensors; equivalent circuits; hydrophones; micromachining; ultrasonic transducer arrays; 2D CMUT array element; EDA environment; acoustic diffraction; acoustic loss; equivalent circuit model; front end IC; hydrophone; pulse echo response; transducer dimension; Acoustic diffraction; Acoustic propagation; Acoustic pulses; Acoustic transducers; Acoustic waves; Equivalent circuits; Integrated circuit modeling; Pulse amplifiers; Pulse circuits; Pulse measurements;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium (IUS), 2009 IEEE International
Conference_Location
Rome
ISSN
1948-5719
Print_ISBN
978-1-4244-4389-5
Electronic_ISBN
1948-5719
Type
conf
DOI
10.1109/ULTSYM.2009.5441944
Filename
5441944
Link To Document