• DocumentCode
    2005989
  • Title

    170 MHz electrodeless quartz crystal microbalance for highly sensitive biosensors

  • Author

    Ogi, Hirotsugu ; Nagai, Hironao ; Fukunishi, Yuji ; Hirao, Masahiko ; Nishiyama, Masayoshi

  • Author_Institution
    Grad. Sch. of Eng. Sci., Osaka Univ., Toyonaka, Japan
  • fYear
    2009
  • fDate
    20-23 Sept. 2009
  • Firstpage
    718
  • Lastpage
    720
  • Abstract
    A highly sensitive wireless-electrodeless quartz-crystal microbalance biosensor with a fundamental resonance frequency of 170 MHz was developed. A blank AT-cut quartz plate of 9.7 ¿m thick was excited by the line antenna in liquid, and its vibration signals were detected by the other line antenna, achieving the noncontacting measurement of the resonance frequency. The mass sensitivity of the 170 MHz QCM biosensor is better than that of a conventional 5 MHz QCM by three orders of magnitude. This biosensor was applied to detect human immunoglobulin G using Staphylococcus aurous protein A immobilized nonspecifically on both surfaces of the quartz plate. The detection limit is 0.5 pM.
  • Keywords
    biosensors; microbalances; quartz; AT-cut quartz plate; SiO2; Staphylococcus aurous protein A; electrodeless quartz crystal microbalance; frequency 170 MHz; highly sensitive biosensors; human immunoglobulin G; mass sensitivity; size 9.7 mum; wireless biosensor; Antenna measurements; Biosensors; Frequency measurement; Humans; Immune system; Resonance; Resonant frequency; Signal detection; Thickness measurement; Vibration measurement; QCM; biosensor; component; electrodeless; noncontadcting measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2009 IEEE International
  • Conference_Location
    Rome
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4244-4389-5
  • Electronic_ISBN
    1948-5719
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2009.5442069
  • Filename