Title :
MEMS fabricated liquid chromatography microchip for practical uses
Author :
Noguchi, Masao ; Tsunoda, Makoto ; Mizuno, Jun ; Funatsu, Takashi ; Shoji, Shuichi
Author_Institution :
Major in Nano-Sci. & Nano-Eng., Waseda Univ., Tokyo, Japan
Abstract :
A microfabricated pillar array separation column which has a comparable theoretical plate number to the conventional packed column was fabricated on a silicon chip. The column has a folded separation microchannel with compensated narrow curve structures which have inner and outer paths of the same length. This geometry minimizes the turn-induced band broadening and achieves high separation efficiency. The liquid chromatography (LC) microchip was fabricated on the 20 mm à 20 mm silicon chip using a column of 110 mm in length. The theoretical plate number is approximately 8000 (typical packed column: 5000-18000) and the plate heights are 10.4 ¿m (typical packed column: 14-20 ¿m). These results show that the microfabricated LC microchip is suitable for practical uses.
Keywords :
chromatography; elemental semiconductors; microfabrication; micromechanical devices; silicon; MEMS; compensated narrow curve structure; folded separation microchannel; liquid chromatography microchip; microfabricated pillar array separation column; silicon chip; size 10.4 micron; size 110 mm; Etching; Geometry; Laboratories; Lithography; Microchannel; Micromechanical devices; Nanotechnology; Pharmaceuticals; Resists; Silicon;
Conference_Titel :
Micro Electro Mechanical Systems (MEMS), 2010 IEEE 23rd International Conference on
Conference_Location :
Wanchai, Hong Kong
Print_ISBN :
978-1-4244-5761-8
Electronic_ISBN :
1084-6999
DOI :
10.1109/MEMSYS.2010.5442355