DocumentCode :
2018657
Title :
Radiation pattern measurement system for millimeter-wave antenna fed by contact probe
Author :
Ito, Takayoshi ; Tsutsumi, Yukako ; Obayashi, Shuichi ; Shoki, Hiroki ; Morooka, Tasuku
Author_Institution :
Toshiba Corp. R&D Center, Kawasaki, Japan
fYear :
2009
fDate :
Sept. 29 2009-Oct. 1 2009
Firstpage :
1543
Lastpage :
1546
Abstract :
This paper presents a radiation pattern measurement system in order to evaluate mm-wave antenna fed by contact probe. It reduces the transmission loss by using a waveguide as a revolving arm. Three cut patterns can be measured by changing the orientation of the revolving arm. A differential feeding dipole antenna with a reflection plate was evaluated by using this measurement system. The measurement result of half-power beam width agrees well with the simulation result and the dynamic range is more than 20 dB. By covering the possible scatterers in the anechoic chamber by planar electromagnetic absorber, the ripple in the half-power beam width is suppressed below plusmn0.5 dB.
Keywords :
anechoic chambers (electromagnetic); antenna radiation patterns; dipole antennas; millimetre wave antennas; anechoic chamber; contact probe; differential feeding dipole antenna; half-power beam width; millimeter-wave antenna fed; planar electromagnetic absorber; radiation pattern measurement; reflection plate; revolving arm; transmission loss; waveguide; Antenna measurements; Antenna radiation patterns; Dipole antennas; Electromagnetic measurements; Electromagnetic scattering; Electromagnetic waveguides; Millimeter wave measurements; Probes; Propagation losses; Reflector antennas;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Conference, 2009. EuMC 2009. European
Conference_Location :
Rome
Print_ISBN :
978-1-4244-4748-0
Type :
conf
Filename :
5296180