  • Title

    Enhanced software based detection of implanted cardiac pacemaker stimuli

  • Author

    Jennings, M. ; Devine, B. ; Luo, S. ; Macfarlane, PW

  • Author_Institution
    Univ. of Glasgow, Glasgow, UK
  • fYear
    2009
  • fDate
    13-16 Sept. 2009
  • Firstpage
    833
  • Lastpage
    836
  • Abstract
    With increasing use and sophistication of implantable cardiac pacemaker devices, new techniques are required to ensure accurate detection of pacemaker stimuli. This study aimed to (i) compare the accuracy of enhanced pacemaker spike detection logic using 32,000 samples/sec (sps) data against existing software based on current AAMI/IEC standards; (ii) assess the ability of the new logic to aid in the detection of biventricular (BiV) pacing; (iii) develop and test a method for the detection of BiV pacing based on QRS morphology. 72 patients were recruited. 63 were used to assess spike detection accuracy and 62 to assess the accuracy of reporting BiV pacing. 33/62 had BiV pacemakers. 9 patients were excluded for various reasons. The new logic improved accuracy of spike detection from 46/63 to 62/63. For BiV pacing, the new logic had a sensitivity of 97% and a specificity of 100% while the QRS morphology logic had a sensitivity of 70% and a specificity of 93%.
  • Keywords
    cardiology; medical signal detection; pacemakers; BiV pacing; QRS morphology logic; biventricular pacing detection; enhanced pacemaker spike detection logic; implanted cardiac pacemaker stimuli; Electrocardiography; Heart; Logic design; Logic devices; Logic testing; Medical treatment; Morphology; Pacemakers; Sampling methods; Sensitivity and specificity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 2009
  • Conference_Location
    Park City, UT
  • ISSN
    0276-6547
  • Print_ISBN
    978-1-4244-7281-9
  • Electronic_ISBN
    0276-6547
  • Type

    conf

  • Filename
    5445253