• DocumentCode
    2038380
  • Title

    The design and implementation of blood analyzer control system based on the fluid flow aperture device

  • Author

    Ling Tao ; Liang Luo ; Fugui Li

  • Author_Institution
    Sch. of Inf. Eng., Nanchang Univ., Nanchang, China
  • fYear
    2015
  • fDate
    2-5 Aug. 2015
  • Firstpage
    174
  • Lastpage
    178
  • Abstract
    Introduce a hardware design and implementation of blood analyzer control system based on the fluid flow aperture device. Control system use 32-bit RISC processors S3C2410A as the core of the electrical system, including driver board, actuators and sensors, input/output devices, etc. The design of special subdivided driving circuit according to the characteristics of the stepping motor ensure the accuracy of the motor, improve the machine performance. At the same time, the control system has designed a unique signal amplifier circuit according to the problems of blood cells in weak signal and complex interference source. It has effectively restrained various disturbance and has ensured accuracy and repeatability of testing results of instrument. It has also provided products of good quality and reasonable price for the domestic numerous small and medium-sized hospitals, and promoted construction of the grassroots township hospitals.
  • Keywords
    amplifiers; biomedical electronics; biomedical equipment; biomedical measurement; blood; cellular biophysics; fluidic devices; medical control systems; reduced instruction set computing; stepping motors; 32-bit RISC processor; S3C2410A; actuator; blood analyzer control system design; blood analyzer control system implementation; blood cell; complex interference source; domestic medium-sized hospital; domestic small-sized hospital; driver board; electrical system; fluid flow aperture device; grassroots township hospital; hardware design; input device; machine performance; motor accuracy; output device; price; quality; sensor; signal amplifier circuit; special subdivided driving circuit design; stepping motor characteristics; testing result accuracy; testing result repeatability; weak signal; Accuracy; Blood; Cells (biology); Control systems; Hardware; Instruments; Liquids; ARM9; Blood analyzer; S3C2410;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Mechatronics and Automation (ICMA), 2015 IEEE International Conference on
  • Conference_Location
    Beijing
  • Print_ISBN
    978-1-4799-7097-1
  • Type

    conf

  • DOI
    10.1109/ICMA.2015.7237477
  • Filename
    7237477