Title :
Design of bioimpedance monitor and its application to atrioventricular delay optimization
Author :
Vondra, Vlastimil ; Viscor, I. ; Halamek, J. ; Jurak, P.
Author_Institution :
Inst. of Sci. Instrum., ASCR, Brno, Czech Republic
Abstract :
We introduce our conception of a two-channel bioimpedance monitor for the impedance cardiography. We are describing the design of a homemade device which is based on a direct digital synthesizer and a digital down converter. This solution enables us to obtain a high-quality bioimpedance signal, which can be used for computing the cardiac output beat per beat. Our device allows simultaneous measuring of the complex impedance in two arbitrary directions on a human/animal body without any interference between the channels. We tested our device during a study of measuring the cardiac output at patients with implanted pacemakers. The cardiac output is measured with respect to the atrio-ventricular delay. We have scored the maximum relative cardiac output for each patient. The result of the study of 26 patients is presented in this paper.
Keywords :
bioelectric phenomena; biomedical measurement; cardiovascular system; electrocardiography; pacemakers; patient monitoring; atrioventricular delay optimization; digital down converter; direct digital synthesizer; impedance cardiography; pacemakers; two-channel bioimpedance monitor; Animals; Bioimpedance; Cardiography; Delay; Design optimization; Humans; Impedance measurement; Interference; Monitoring; Synthesizers;
Conference_Titel :
Computers in Cardiology, 2009
Conference_Location :
Park City, UT
Print_ISBN :
978-1-4244-7281-9
Electronic_ISBN :
0276-6547