Abstract :
The article contains further development of the author´s problem. I have tested the application of the ionization chamber as a dosimeter. I was measuring the changes of ionization current with the absorption of X-rays by metals. I was changing the thickness of the absorbent for different voltage of the X-lamp. The obtained results have the same character as in the case of other dosimeters used so far (e.g. Geiger-Mueller counter). Therefore my ionization chamber can function as dosimetric chamber (e.g. in medicine), although further research would be demanded to ensure better usage.
Keywords :
X-ray absorption; X-ray detection; dielectric liquids; dosimeters; liquid ionisation chambers; liquid mixtures; X-ray absorption; dosimeter; ionization chamber; ionization current; liquid dielectric mixtures; medicine; metals; Current measurement; Dielectric liquids; Dielectric measurements; Electromagnetic wave absorption; Ionization chambers; Silicon; Temperature; Testing; Voltage; X-rays;