• DocumentCode
    2039555
  • Title

    A CMOS turbidity to frequency converter with calibration circuits for detecting turbidity applications

  • Author

    Chiang, Cheng-Ta ; Huang, Shih-Ming

  • Author_Institution
    Department of Electrical Engineering, National Chia Yi University, Taiwan
  • fYear
    2015
  • fDate
    2-5 Aug. 2015
  • Firstpage
    381
  • Lastpage
    385
  • Abstract
    In this paper, a CMOS turbidity to frequency converter with calibration circuits for detecting turbidity applications is newly proposed. A calibration technique is proposed in this work. Another innovation is that the outputs of the proposed chip are directly digitized, they could be easily sent over a wide range of transmission media, such as PSN, radio, optical, IR, ultrasonic, and etc. Before performing the proposed calibration circuits, the maximum linear error is 25.82%. After calibration, the maximum linear error is reduced to 0.89%. The turbidity ranges from 1.5 to 992 Nephelometric Turbidity Units (NTU), and its corresponding output frequency range is 24.875 to 18.214 kHz. The chip area is 1.01 × 1.01 mm2. The proposed chip is suitable for devices of measuring water turbidity.
  • Keywords
    CMOS integrated circuits; Calibration; Capacitance; Frequency conversion; Optical sensors; Robot sensing systems; Simulation; calibration circuits; turbidity sensor; water turbidity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Mechatronics and Automation (ICMA), 2015 IEEE International Conference on
  • Conference_Location
    Beijing, China
  • Print_ISBN
    978-1-4799-7097-1
  • Type

    conf

  • DOI
    10.1109/ICMA.2015.7237515
  • Filename
    7237515