A cardiovascular model for the analysis of pacing configurations in cardiac resynchronization therapy
Author :
Ve Koon, K. Tse ; Le Rolle, V. ; Carrault, G. ; Hernández, A.I.
Author_Institution :
LTSI, Univ. de Rennes 1, Rennes, France
fYear :
2009
fDate :
13-16 Sept. 2009
Firstpage :
393
Lastpage :
396
Abstract :
A lumped-parameter model of the cardiovascular system, made of 3 interacting modules (an electrical heart model, an elastance-based mechanical heart model and a circulatory system) is proposed in this paper. This model is able to reproduce the clinically observed quasi-parabolic nature of the dependence of systolic blood pressure (SBP) on atrioventricular (AVD) and interventricular (VVD) delays and the increased SBP on AVD dependance at high heart rates. A sensitivity analysis has been performed on the main variables of this model, as well as parameter identification on interpolated clinical data. Results from these analyses show the importance of integrating the circulatory and regulatory systems for the reproduction of appropriate physiological responses to CRT.
Keywords :
bioelectric phenomena; cardiovascular system; lumped parameter networks; pacemakers; parameter estimation; physiological models; AVD; SBP; VVD; atrioventricular delay; cardiac resynchronization therapy; cardiovascular model; interventricular delay; lumped-parameter model; pacing configurations; parameter identification; sensitivity analysis; systolic blood pressure; Blood pressure; Cardiology; Cardiovascular system; Cathode ray tubes; Circulatory system; Delay; Heart rate; Medical treatment; Parameter estimation; Sensitivity analysis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology, 2009
Conference_Location :
Park City, UT
ISSN :
0276-6547
Print_ISBN :
978-1-4244-7281-9
Electronic_ISBN :
0276-6547
Type :
conf
Filename :
5445386
Link To Document :
بازگشت