    Verification of Set Up Dual-Chamber Pacemaker Electrical Parameters

  • Author

    Augustynek, Martin ; Labza, Z. ; Penhaker, Marek ; Korpas, D.

  • Author_Institution
    Dept. of Meas. & Control, VSB - Tech. Univ. of Ostrava, Ostrava, Czech Republic
  • Volume
    2
  • fYear
    2010
  • fDate
    19-21 March 2010
  • Firstpage
    168
  • Lastpage
    172
  • Abstract
    The measurement on a dual-chamber cardiostimulator consists of the construction of a unit for the simulation of the cardiac action in a patient´s body. It has an adjustable amplitude (0 - 20 mV), frequency (0 - 100 Hz) and mark-space ratio. The system output impedance should be in the range of 200 - 2000 ¿. It is possible to add a defined noise into the output channel. It is further possible to measure the tolerance of sensing of the atrial and ventricular channel of the dual-chamber cardiostimulator DDDR.
  • Keywords
    bioelectric phenomena; cardiology; pacemakers; atrial channel; dual-chamber cardiostimulator; dual-chamber pacemaker electrical parameters; set up verification; ventricular channel; Cardiology; Circuit simulation; Electric variables measurement; Electrodes; Frequency; Impedance; Mathematical model; Pacemakers; Printed circuits; Voltage; detection; measurement methods; stimulation voltage; transform cardiostimulator;