• DocumentCode
    2050884
  • Title

    Penning effects in Ar-Xe mixtures at atmospheric pressures: absolute W-value measurements for 5.9 keV X-rays

  • Author

    Vinagre, F.L.R. ; Conde, C.A.N.

  • Author_Institution
    Dept. de Fisica, Coimbra Univ., Portugal
  • Volume
    2
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    651
  • Abstract
    Absolute measurements of W-values for 5.9 keV X-rays in Ar-Xe mixtures were obtained using an experimental method recently developed, based on the collection of the primary ionization charge and the counting of the number of photons absorbed. The absolute W-values obtained for the following Ar-Xe mixtures at 825 Torr, 10%, 20% and 40% Xe, are 21.42, 21.64, 21.63 eV, respectively, showing the influence of Penning effects in the ionization yields of the mixtures. The accuracy achieved in these measurements is ±1%. The benefits and drawbacks of these mixtures as filling gases for gas proportional scintillation counters are discussed
  • Keywords
    Penning ionisation; X-ray detection; argon; gas mixtures; gas scintillation detectors; proportional counters; xenon; 5.9 keV; 825 torr; Ar-Xe; Ar-Xe mixtures; Penning effects; absolute W-value measurements; atmospheric pressures; gas proportional scintillation counters; ionization yields; primary ionization charge; Atmospheric measurements; Charge measurement; Current measurement; Electrons; Ionization; Ionizing radiation; Photonic integrated circuits; Physics; Scintillation counters; X-rays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.845753
  • Filename
    845753