Title :
Improved Calibration Method for Lateral Force of the Cantilever Deflection Force Sensor in Atomic Force Microscope
Author :
Wang, Fei ; Zhao, Xuezeng
Author_Institution :
Sch. of Mech. & Electron. Eng., Harbin Inst. of Technol.
Abstract :
Quantitative friction measurement of nano materials in atomic force microscope (AFM) requires accurate calibration method for the lateral force. An improved calibration method for the lateral force of the cantilever in AFM is presented. The calibration factor derived from the original method increased with the applied normal load, which indicates that separate calibration should be required for every given applied normal load to keep the accuracy of friction measurement. We analyze the reason for this phenomenon, and improve the original method by introducing the contact factor, which is derived from the contact stiffness between the tip and the sample, to the calculation of calibration factors. Our method makes the calculation of calibration factors under different applied normal loads be possible without repeating the calibration procedure. Comparative experiments on a silicon wafer have been done by both the two methods to validate our method
Keywords :
atomic force microscopy; calibration; force measurement; force sensors; applied normal load; atomic force microscope; calibration factor; cantilever deflection force sensor; contact factor; contact stiffness; lateral force; nanomaterials; quantitative friction measurement; silicon wafer; Atomic force microscopy; Atomic measurements; Calibration; Electron microscopy; Error correction; Force measurement; Force sensors; Friction; Systems engineering and theory; Torque measurement; atomic force microscope; calibration; cantilever;
Conference_Titel :
Nano/Micro Engineered and Molecular Systems, 2006. NEMS '06. 1st IEEE International Conference on
Conference_Location :
Zhuhai
Print_ISBN :
1-4244-0139-9
Electronic_ISBN :
1-4244-0140-2
DOI :
10.1109/NEMS.2006.334616