• DocumentCode
    2052897
  • Title

    Image reconstruction from misaligned truncated helical cone-beam data

  • Author

    Noo, E. ; Schaller, S. ; Clackdoyle, R. ; White, T.A. ; Roney, T.J.

  • Author_Institution
    Dept. of Appl. Math., Liege Univ., Belgium
  • Volume
    2
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    1028
  • Abstract
    The effects of a misalignment between the rotation and translation axes of the scanner in helical cone-beam CT was investigated. A simple method to estimate the misalignment parameters was proposed and the PHI-method was shown to be suitable to obtain accurate results from misaligned data
  • Keywords
    Radon transforms; computerised tomography; diagnostic radiography; image reconstruction; medical image processing; DAQ geometry; PHI-method; Radon data; helical cone-beam CT; image reconstruction; long object problem; misaligned truncated helical cone-beam data; misalignment parameters; rebinning; rotation axes; translation axes; Biomedical imaging; Computed tomography; Data acquisition; Detectors; Geometry; Image reconstruction; Mathematics; Object detection; Radiology; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.845837
  • Filename