• DocumentCode
    2053028
  • Title

    Design of a depth of interaction detector with a PS-PMT for PET

  • Author

    Murayama, H. ; Ishibashi, H. ; Uchida, H. ; Omura, T. ; Yamashita, T.

  • Author_Institution
    Nat. Inst. of Radiol. Sci., Chiba, Japan
  • Volume
    2
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    1062
  • Abstract
    A detector unit using a depth encoding scheme was designed and evaluated. The detector unit consists of four crystal blocks in a 2×2 array coupled to a position-sensitive photomultiplier tube (PS-PMT) having metal channel dynodes and 4×4 multianodes. Each crystal block consists of three stages of 2×2 GSO arrays of which element size is 3.8 mm×3.8 mm×10 mm. Each scintillation event is mapped on a two dimensional positioning histogram according to the relative ratio of the output signals of the PS-PMT. With appropriate light control in the crystal block, good positioning performance for Cs-137 photons was obtained
  • Keywords
    biomedical equipment; gamma-ray detection; image coding; medical image processing; photomultipliers; positron emission tomography; Cs; Cs-137 photons; PET; PS-PMT; crystal blocks; depth encoding scheme; depth of interaction detector design; element size; light control; medical diagnostic imaging; metal channel dynodes; multianodes; nuclear medicine; position-sensitive photomultiplier tube; positioning performance; two dimensional positioning histogram; Gamma rays; Image resolution; Lighting control; Optical arrays; Photomultipliers; Photonic crystals; Position sensitive particle detectors; Positron emission tomography; Sensor arrays; Solid scintillation detectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.845844
  • Filename
    845844