Title :
Evaluation of a Calorimetric Thermal Voltage Converter for RF–DC Difference up to 1 GHz
Author :
Yueyan Shan ; Yu Song Meng ; Filipski, Piotr S.
Author_Institution :
Agency for Sci., Technol. & Res., Nat. Metrol. Centre, Singapore, Singapore
Abstract :
In this paper, we present a performance evaluation of the RF-DC transfer difference for a calorimetric thermal voltage converter (CTVC) designed by the National Research Council of Canada (NRC) at frequencies up to 1 GHz. In the first part of this paper, we describe a bilateral comparison of the RF-DC difference standards between the NRC and the National Metrology Centre of the Agency for Science, Technology and Research of Singapore, in the frequency band from 1 kHz to 100 MHz. A good agreement is observed between the two laboratories using the CTVC as a traveling standard. In the second part of this paper, we evaluate the performance of the CTVC at higher frequencies up to 1 GHz. In this part, RF-DC difference of the CTVC is mathematically modeled and experimentally evaluated in terms of the calibration factor of a thermistor mount and the reflection coefficients at its type-N input connector.
Keywords :
calibration; calorimeters; calorimetry; radiofrequency measurement; thermoelectricity; transfer standards; voltage measurement; Agency for Science Technology and Research of Singapore; CTVC; NRC; National Metrology Centre; National Research Council of Canada; RF-DC transfer difference; calorimetric thermal voltage converter; Calibration; Connectors; Mathematical model; Radio frequency; Standards; Thermistors; Uncertainty; ACndash;DC difference; RF–DC difference; calorimetric thermal voltage converter (CTVC); mathematical modeling;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2013.2278597