• DocumentCode
    2061192
  • Title

    The infrared thermal converter as a highly precise transfer standard

  • Author

    Castelli, Franco

  • Author_Institution
    Dipartimento di Elettrotecnica, Politecnico di Milano, Italy
  • Volume
    1
  • fYear
    1996
  • fDate
    1996
  • Firstpage
    206
  • Abstract
    The new thermal converter developed has the following AC-DC transfer errors: lower than 2.5·10-4 ppm (1 σ) at frequencies from DC to 75 kHz; equal to: 10-3 ppm 100 kHz and to 0.01 ppm at 125 kHz. The design is based on an infrared thermal sensor irradiated by a “flat strap sandwich heater resistor” (FSSHR). The high accuracy results qualify the converter as a possible AC-DC reference transfer standard
  • Keywords
    calibration; current comparators; electric current measurement; infrared detectors; measurement standards; temperature sensors; voltage measurement; 100 kHz; 125 kHz; AC-DC transfer errors; calibration; current comparator; flat strap sandwich heater resistor; highly precise transfer standard; infrared thermal converter; infrared thermal sensor; reference transfer standard; shunt; thermal loss; Analog-digital conversion; Frequency conversion; Infrared heating; Infrared sensors; Measurement standards; Resistors; Tellurium; Testing; Thermal resistance; Thermal sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1996. IMTC-96. Conference Proceedings. Quality Measurements: The Indispensable Bridge between Theory and Reality., IEEE
  • Conference_Location
    Brussels
  • Print_ISBN
    0-7803-3312-8
  • Type

    conf

  • DOI
    10.1109/IMTC.1996.507377
  • Filename
    507377