SAR value variation by ambient temperature of the SAR measurement system
Author
Yoon Myoung Gimm
Author_Institution
Sch. of Electron. & Electr. Eng., Dankook Univ., Yongin
fYear
2008
fDate
19-23 May 2008
Firstpage
124
Lastpage
127
Abstract
Validating SAR values change by the ambient environmental temperature of the test room. Measured SAR value variations by the E-field probe response to the air temperature and by the electrical parameters of the human simulant liquid are analysed in this study. The prompt input of the electrical parameters of the liquid to the measurement system is important to measure SAR correctly.
Keywords
absorption; biological effects of fields; E-field probe response; SAR measurement system; SAR value variation; ambient environmental temperature; human simulant liquid; Antenna measurements; Electric variables measurement; Electromagnetic waveguides; Frequency; Humans; Imaging phantoms; Permittivity measurement; Probes; Temperature distribution; Temperature measurement;