• DocumentCode
    2069321
  • Title

    SAR value variation by ambient temperature of the SAR measurement system

  • Author

    Yoon Myoung Gimm

  • Author_Institution
    Sch. of Electron. & Electr. Eng., Dankook Univ., Yongin
  • fYear
    2008
  • fDate
    19-23 May 2008
  • Firstpage
    124
  • Lastpage
    127
  • Abstract
    Validating SAR values change by the ambient environmental temperature of the test room. Measured SAR value variations by the E-field probe response to the air temperature and by the electrical parameters of the human simulant liquid are analysed in this study. The prompt input of the electrical parameters of the liquid to the measurement system is important to measure SAR correctly.
  • Keywords
    absorption; biological effects of fields; E-field probe response; SAR measurement system; SAR value variation; ambient environmental temperature; human simulant liquid; Antenna measurements; Electric variables measurement; Electromagnetic waveguides; Frequency; Humans; Imaging phantoms; Permittivity measurement; Probes; Temperature distribution; Temperature measurement;
  • fLanguage
    English