DocumentCode
2072976
Title
A Signal Processing System Used in Digital Ophthalmology Ultrasonography
Author
Zhou, Sheng ; Wang, Xiaochun ; Wang, Yanqun ; Yang, Jun
Author_Institution
Biomed. Eng. Inst., Chinese Acad. of Med. Sci., Tianjin, China
fYear
2009
fDate
17-19 Oct. 2009
Firstpage
1
Lastpage
5
Abstract
Objective to develop a signal processing system based on FPGA used in digital ophthalmology ultrasongraphy to process echo signals. Methods main signal processing technique used in the system includes interpolation, dynamic filter, TGC, envelop demodulation and logarithmic amplifier. In FPGA, a schematic file acts as the top module, and hardware describe language VHDL act as bottom modules. Results By building a physical simulation model, the validity of each phases in signal processing of the system has been validated. By detecting normal human eyes and orbits, the echo signals gained are all right. Conclusion The system has good ability to process high frequency ultrasonic echo signals of ophthalmology in real time. It has reached design demands, and shows a good application prospect.
Keywords
amplifiers; biomedical ultrasonics; demodulation; echo; eye; field programmable gate arrays; filtering theory; hardware description languages; interpolation; medical signal detection; FPGA; TGC; VHDL; digital ophthalmology ultrasonography; dynamic filtering; echo signal processing system; envelop demodulation; hardware describe language; human eye detection; interpolation; logarithmic amplifier; orbits; schematic file; Demodulation; Digital signal processing; Eyes; Field programmable gate arrays; Filters; Hardware; Humans; Interpolation; Signal processing; Ultrasonography;
fLanguage
English
Publisher
ieee
Conference_Titel
Image and Signal Processing, 2009. CISP '09. 2nd International Congress on
Conference_Location
Tianjin
Print_ISBN
978-1-4244-4129-7
Electronic_ISBN
978-1-4244-4131-0
Type
conf
DOI
10.1109/CISP.2009.5301038
Filename
5301038
Link To Document