Network Analyzer Calibration Without Impedance Standards
Author
Jöstingmeier, A. ; Faby, G. ; Jenett, M. ; Schunemann, Klaus
Author_Institution
Technische Universitÿt Hamburg-Harburg
Volume
1
fYear
1990
fDate
9-13 Sept. 1990
Firstpage
203
Lastpage
208
Abstract
This paper presents an alternative method for network analyzer calibration which shows three advantages: It avoids de-embedding the device under test, it does not need any impedance standard, and it allows characterizing an unknown two-port behind an arbitrary number of cascaded two-ports. It is thus directly applicable to planar circuits and to MICs.
Keywords
Calibration; Circuit testing; Impedance; Measurement errors; Millimeter wave measurements; Nonlinear equations; Reflection; Scattering parameters; Transmission line measurements; Wavelength measurement;