The comparison of a novel continuous cardiac output monitor based on pulse wave transit time and echo Doppler during exercise
Author :
Sugo, Y. ; Sakai, Tadashi ; Terao, M. ; Ukawa, T. ; Ochiai, R.
Author_Institution :
Nihon Kohden Corp., Tokyo, Japan
fYear :
2012
fDate :
Aug. 28 2012-Sept. 1 2012
Firstpage :
236
Lastpage :
239
Abstract :
A new technology called estimated continuous cardiac output (esCCO) uses pulse wave transit time (PWTT) obtained from an electrocardiogram and pulse oximeter to measure cardiac output (CO) non-invasively and continuously. This study was performed to evaluate the accuracy of esCCO during exercise testing. We compared esCCO with CO measured by the echo Doppler aortic velocity-time integral (VTIao_CO). The correlation coefficient between esCCO and VTIao_CO was r= 0.87 (n= 72). Bias and precision were 0.33 ± 0.95 L/min and percentage error was 31%. The esCCO could detect change in VTIao_CO larger than 1 L/min with a concordance rate of 88%. In polar plot, 83% of data are within 0.5 L/min, and 100% of data are within 1 L/min. Those results show the acceptable accuracy and trend ability of esCCO. Change in pre-ejection period (PEP) measured by using M-mode of Diagnostic Ultrasound System accounted for approximately half of change in PWTT. This indicates that PEP included in PWTT has an impact on the accuracy of esCCO measurement. In this study, the validity of esCCO during exercise testing was assessed and shown to be acceptable. The result of this study suggests that we can expand its application.
Keywords :
Doppler measurement; echocardiography; electrocardiography; oximetry; patient monitoring; M-mode diagnostic ultrasound system; cardiac output measurement; correlation coefficient; echo Doppler aortic velocity-time integral; electrocardiogram; estimated continuous cardiac output; exercise testing; novel continuous cardiac output monitoring; pulse oximetry; pulse wave transit time; Biomedical monitoring; Doppler effect; Electrocardiography; Monitoring; Protocols; Testing; Ultrasonic variables measurement; Adult; Cardiac Output; Echocardiography, Doppler; Electrocardiography; Exercise Test; Humans; Male; Oximetry; Pulse Wave Analysis; Sensitivity and Specificity;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location :
San Diego, CA
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4119-8
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2012.6345913
Filename :
6345913
Link To Document :
بازگشت