Abstract :
Cardiovascular disease caused 32.8% of deaths in the United States in 2008 [1]. The most important medical parameter is the arterial blood pressure. The origin of high or low blood pressure can mostly be found in the vessel compliance. With the presented implantable sensor, we are able to directly measure strain of arteries, as an indicator of arteriosclerosis. The sensor is designed as a cuff with integrated capacitive structures and is wrapped around arteries. With a new and innovative locking method, we could show that the system does not affect the arteries. This is demonstrated by theory as well as experimental in vivo investigations. Biocompatibility tests, confirmed by histological cuts and MRI measurements, showed that no stenosis, allergic reactions or inflammation occurs. The sensor shows excellent linear behavior with respect to stress and strain.
Keywords :
biomechanics; biomedical measurement; blood vessels; capacitive sensors; haemodynamics; prosthetics; strain measurement; arterial strain measurement; arteriosclerosis indicator; cardiovascular disease; cuff sensor; high blood pressure; implantable capacitive sensor; integrated capacitive structures; low blood pressure; vessel compliance; Arteries; Capacitance; Magnetic resonance imaging; Materials; Strain; Strain measurement; Stress; Animals; Arterial Pressure; Arteries; Arteriosclerosis; Blood Pressure Determination; Compliance; Equipment Design; Femoral Artery; Humans; Magnetic Resonance Angiography; Models, Cardiovascular; Prostheses and Implants; Silicones; Sus scrofa; Vascular Capacitance;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location :
San Diego, CA
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4119-8
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2012.6345986
Filename :
6345986
Link To Document :
بازگشت