    Detection of pressure and temperature about gallstone auto-dissolving instrument

  • Author

    Wang, Yuan-sheng ; Kang, Hong-mei ; Lu, Gui-Ying ; Bo, Li

  • Author_Institution
    Dept of Mechanical Engineering, School of Mechanical and Electrical Information, China University of GeoScience (Wuhan), China
  • fYear
    2010
  • fDate
    4-6 Dec. 2010
  • Firstpage
    199
  • Lastpage
    202
  • Abstract
    Directly dissolving gallstone by pouring MBTE in Gallbladder is a valid therapy. Gallstone auto-dissolving instrument replace manual perfusion by auto-perfusion, while the pressure of inner gallbladder and the temperature of MBTE solvent must keep in a rational range, thus MBTE´s dissolving potential can be exerted. A system made up with NORCO-6715 IPC, AC6111 Acquisition board, NPC1210 pressure sensor and BTS101 temperature sensor can ensure a precision of ±0.1kPa and ±1°C respectively. Out-body test shows the design meets the desire.