DocumentCode
2084052
Title
Non-stationary EMF surveying accuracy improvement
Author
Bienkowski, Pawel ; Trzaska, Hubert
Author_Institution
EM Environ. Protection Lab., Tech. Univ. of Wroclaw, Wroclaw, Poland
fYear
2009
fDate
16-20 Sept. 2009
Firstpage
289
Lastpage
292
Abstract
The paper discusses accuracy of non-stationary EMF measurements. The accuracy is mainly limited by applied methods of meters calibration. New method of calibration is presented. The method allows a limitation of the calibration inaccuracy to level of plusmn1 dB.
Keywords
calibration; electromagnetic field theory; accuracy improvement; calibration inaccuracy; electromagnetic field; nonstationary EMF measurements; nonstationary EMF surveying; Computer architecture; Formal specifications; Internet; Load management; Petri nets; Quality of service; Resource management; Service oriented architecture; Web server; Workstations;
fLanguage
English
Publisher
ieee
Conference_Titel
Environmental Electromagnetics, 2009. CEEM 2009. 5th Asia-Pacific Conference on
Conference_Location
Xian
Print_ISBN
978-1-4244-4344-4
Type
conf
DOI
10.1109/CEEM.2009.5301445
Filename
5301445
Link To Document