DocumentCode :
2086806
Title :
Performance evaluation of a lower limb exoskeleton for stair ascent and descent with Paraplegia
Author :
Farris, R.J. ; Quintero, H.A. ; Goldfarb, Michael
Author_Institution :
Vanderbilt Univ., Nashville, TN, USA
fYear :
2012
fDate :
Aug. 28 2012-Sept. 1 2012
Firstpage :
1908
Lastpage :
1911
Abstract :
This paper describes the application of a powered lower limb exoskeleton to aid paraplegic individuals in stair ascent and descent. A brief description of the exoskeleton hardware is provided along with an explanation of the control methodology implemented to allow stair ascent and descent. Tests were performed with a paraplegic individual (T10 complete injury level) and data is presented from multiple trials, including the hip and knee joint torque and power required to perform this functionality. Joint torque and power requirements are summarized, including peak hip and knee joint torque requirements of 0.75 Nm/kg and 0.87 Nm/kg, respectively, and peak hip and knee joint power requirements of approximately 0.65 W/kg and 0.85 W/kg, respectively.
Keywords :
injuries; prosthetics; Paraplegia; T10 complete injury level; hip torque; knee joint torque; lower limb exoskeleton; performance evaluation; stair ascent; stair descent; Exoskeletons; Foot; Hip; Joints; Knee; Legged locomotion; Torque; Body Weight; Evaluation Studies as Topic; Humans; Lower Extremity; Male; Orthotic Devices; Paraplegia; Posture; Torque; Walking;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location :
San Diego, CA
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4119-8
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2012.6346326
Filename :
6346326
Link To Document :
بازگشت