Abstract :
Low field MRI machines are important, since the machines offer easy access for users. However, the SNR attained with a low field MRI machine is not necessarily high. This paper attempts to present an effective resonator coil designed in particular for a human hand placed in a wrist-sized permanent magnet MRI machine. The optimal design of a solenoidal coil is investigated. The wire conductor features and the proximity to the other conductors are also investigated. A rectangular solenoidal coil was built. MR Images of the hand were obtained. The wrist contour, finger muscles, fatty tissues and bone marrow were distinctly displayed.
Keywords :
biomedical MRI; biomedical equipment; bone; muscle; permanent magnets; resonator filters; solenoids; tissue engineering; bone marrow; effective resonator coil; fatty tissues; finger muscles; magnetic flux density 0.2 T; rectangular solenoidal coil; wrist contour; wrist-sized permanent magnet MRI machine; Bones; Coils; Conductors; Fingers; Humans; Magnetic resonance imaging; Muscles; Permanent magnets; Wire; Wrist;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Complex Medical Engineering, 2007. CME 2007. IEEE/ICME International Conference on
Conference_Location :
Beijing
Print_ISBN :
978-1-4244-1077-4
Electronic_ISBN :
978-1-4244-1078-1
Type :
conf
DOI :
10.1109/ICCME.2007.4381863
Filename :
4381863
Link To Document :
بازگشت