A digital instrument for venous muscle pump test
Author
Bhooma, G. ; Kokila, S. ; Jayanthi, K. ; Kumar, Varadarajan Jagadeesh
Author_Institution
Dept. of Electron. & Commun. Eng., Pondicherry Eng. Coll., Pondicherry, India
fYear
2011
fDate
10-12 May 2011
Firstpage
1
Lastpage
4
Abstract
Design and development of a prototype unit to measure venous refilling time (VRT) and muscle pump capacity (MPC) during a venous muscle pump test (VMPT) is presented here. Instead of normalizing the photo-plethysmo-graph (PPG), as is common in the VMPT instruments in vogue, a suitable signal processing technique that extracts a signal proportional only to the blood volume changes underneath the PPG sensor is employed. Apart from the measurement of VRT and MPC, the prototype unit also provides a measure on the pulse rate.
Keywords
biomedical equipment; blood pressure measurement; blood vessels; muscle; photoplethysmography; PPG sensor; blood volume changes; digital instrument; muscle pump capacity; photoplethysmo-graph; prototype unit; pulse rate measurement; signal processing technique; signal proportionality; venous muscle pump test; venous refilling time; Blood; Detectors; Muscles; Prototypes; Pulse measurements; Software; Veins; Biomedical measurements; Biomedical transducers; Optical sensors; Photoplethysmography; Venous refilling time; microcontroller; muscle pump capacity;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference (I2MTC), 2011 IEEE
Conference_Location
Binjiang
ISSN
1091-5281
Print_ISBN
978-1-4244-7933-7
Type
conf
DOI
10.1109/IMTC.2011.5944063
Filename
5944063
Link To Document