Title :
Update of two dimensional arrays for real time volumetric and real time intracardiac imaging
Author :
Light, Edward D. ; Fiering, Jason O. ; Hultman, Peter A. ; Lee, Warren ; Smith, Stephen W.
Author_Institution :
Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
Abstract :
The authors have previously described 2-D array transducers with up to several thousand elements operating at frequencies between 2.5 and 5.0 MHz for real time volumetric imaging. Lately, there has been interest in developing catheter based intracardiac imaging systems to aid in the precise tracking of anatomical features and intracardiac devices for improved diagnoses and therapies. The authors constructed several arrays for real time intracardiac volumetric imaging based upon 2 different designs; a 13×11=143 element 5.0 MHz 2-D array for side scanning applications, and a 10×10=100 element 7.0 MHz 2-D array for side scanning applications. The 5.0 MHz array fits into a 12 French (3.8 mm OD) catheter and the 7.0 MHz transducer is designed to fit into a 9 French (2.9 mm OD) catheter. The authors also constructed 2 transducers for transthoracic imaging; a 40×40 5.0 MHz 2-D array, and a 40×40 7.0 MHz 2-D array. The -6 dB fractional bandwidths for the transducers varied from 27% to 67%. All the transducers were constructed on a 6 layer polyimide interconnect. Both transthoracic and intracardiac volumetric images of ultrasound phantoms and animal models have been obtained using the Duke University real time volumetric imaging system which is capable of generating multiple planes at any desired angle and depth within a pyramidal volume
Keywords :
biomedical transducers; biomedical ultrasonics; cardiology; ultrasonic transducer arrays; 2.5 to 7.0 MHz; 2.9 mm; 3.8 mm; 6 layer polyimide interconnect; Duke University; anatomical features; anatomical features tracking; intracardiac devices; medical diagnostic imaging; multiple planes generation; pyramidal volume; real time intracardiac imaging; real time volumetric imaging; side scanning applications; transthoracic imaging; two dimensional arrays; Animals; Bandwidth; Catheters; Frequency; Imaging phantoms; Medical treatment; Polyimides; Real time systems; Ultrasonic imaging; Ultrasonic transducers;
Conference_Titel :
Ultrasonics Symposium, 1999. Proceedings. 1999 IEEE
Conference_Location :
Caesars Tahoe, NV
Print_ISBN :
0-7803-5722-1
DOI :
10.1109/ULTSYM.1999.849216