Title :
Experimental study of transmit synthetic focusing combined with receive dynamic focusing in B-mode ultrasound imaging systems
Author :
Bae, Moo-Ho ; Jeong, Mok-Kun ; Song, Tae-Kyong ; Ahn, Young-Bok
Author_Institution :
Medison Co. Ltd., Seoul, South Korea
Abstract :
Recently, a method for focusing at all imaging points both in transmit and receive was proposed. This method uses transmit synthetic focusing and receive dynamic focusing in conventional B-mode ultrasound imaging systems. Transmit focusing is synthesized using the RF data of each scanline, together with all other relevant scanline data previously stored. Experimental results show that the proposed method improves lateral resolution and SNR of real-time B-mode images at all imaging depth points
Keywords :
biomedical ultrasonics; ultrasonic focusing; B-mode ultrasound imaging systems; SNR; imaging depth points; lateral resolution improvement; medical diagnostic imaging; real-time B-mode images; receive dynamic focusing; scanline RF data; transmit synthetic focusing; Bidirectional control; Delay effects; Focusing; Image resolution; Poles and towers; Signal resolution; Signal to noise ratio; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
Conference_Titel :
Ultrasonics Symposium, 1999. Proceedings. 1999 IEEE
Conference_Location :
Caesars Tahoe, NV
Print_ISBN :
0-7803-5722-1
DOI :
10.1109/ULTSYM.1999.849226