• DocumentCode
    2098800
  • Title

    Experimental study of transmit synthetic focusing combined with receive dynamic focusing in B-mode ultrasound imaging systems

  • Author

    Bae, Moo-Ho ; Jeong, Mok-Kun ; Song, Tae-Kyong ; Ahn, Young-Bok

  • Author_Institution
    Medison Co. Ltd., Seoul, South Korea
  • Volume
    2
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    1261
  • Abstract
    Recently, a method for focusing at all imaging points both in transmit and receive was proposed. This method uses transmit synthetic focusing and receive dynamic focusing in conventional B-mode ultrasound imaging systems. Transmit focusing is synthesized using the RF data of each scanline, together with all other relevant scanline data previously stored. Experimental results show that the proposed method improves lateral resolution and SNR of real-time B-mode images at all imaging depth points
  • Keywords
    biomedical ultrasonics; ultrasonic focusing; B-mode ultrasound imaging systems; SNR; imaging depth points; lateral resolution improvement; medical diagnostic imaging; real-time B-mode images; receive dynamic focusing; scanline RF data; transmit synthetic focusing; Bidirectional control; Delay effects; Focusing; Image resolution; Poles and towers; Signal resolution; Signal to noise ratio; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1999. Proceedings. 1999 IEEE
  • Conference_Location
    Caesars Tahoe, NV
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-5722-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1999.849226
  • Filename
    849226