• DocumentCode
    2103486
  • Title

    Electro-Optic-Probe system response: experiment and simulation

  • Author

    Cheng, H.J. ; Whitaker, J.F. ; Herrick, K.J. ; Dib, N. ; Katehi, L.P.B.

  • Volume
    1
  • fYear
    1995
  • fDate
    4-4 Sept. 1995
  • Firstpage
    151
  • Lastpage
    154
  • Abstract
    The frequency response of an extemal electro-optic probe has been determined, experimeritally and using finite-difference, time-domain simulations, to be extremely sensitve to both the elevation and permittivity of the probe. An analytical expression to correct for the system response of the probe has been tested. The use of low-permittivity EO polymers has also been proposed and demonstrated computationally.
  • Keywords
    Calibration; Circuits; Computational modeling; Finite difference methods; Frequency measurement; Power system transients; Probes; Pulse measurements; Time domain analysis; Transmission line measurements;