Title :
The dual mixer time difference at IEN
Author_Institution :
Ist. Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
Abstract :
We report the results obtained with a frequency instability measurement system based upon the dual-mixer time-difference scheme working at 10 MHz with a 1 Hz beat note. It shows a two samples (Allan deviation) measurement floor of 5.6.10/sup -14///spl tau/ and a temperature sensitivity of some picoseconds per Celsius.
Keywords :
Frequency stability; Frequency standards; Mixers (circuits); Time measurement; 10 MHz; Allan deviation; beat note; crosstalk problems; double balanced mixers; dual-mixer time-difference; frequency instability measurement system; frequency standards comparison; measurement floor; temperature sensitivity; zero crossing detector; Counting circuits; Crosstalk; Detectors; Frequency; Jitter; Mixers; Noise measurement; Oscillators; Prototypes; Time measurement;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2000 Conference on
Conference_Location :
Sydney, NSW, Australia
Print_ISBN :
0-7803-5744-2
DOI :
10.1109/CPEM.2000.850872