DocumentCode :
2126838
Title :
International comparison for RF power in the frequency range up to 18 GHz
Author :
de Vreede, J.P.M. ; Korfage-Janssen, W.
Author_Institution :
NMi Van Swinden Lab., Delft, Netherlands
fYear :
2000
fDate :
14-19 May 2000
Firstpage :
58
Lastpage :
59
Abstract :
A Euromet international laboratory comparison (Project 393) has been carried out between 15 national standards laboratories. Thermistor mounts were used, equipped with PC7 as well as Type N connectors. The results show good agreement in measuring the calibration coefficient of the DUTs and in measuring their reflection coefficients.
Keywords :
Calibration; Measurement uncertainty; Microwave measurement; Power measurement; Thermistors; Transfer standards; UHF measurement; 10 MHz to 18 GHz; Euromet project; RF power measurement; calibration coefficient; international laboratory comparison; national standards laboratories; reflection coefficients; thermistor mounts; travelling scheme; uncertainties; Calibration; Connectors; Hafnium; Laboratories; Metrology; Particle measurements; Radio frequency; Reflection; Telecommunications; Thermistors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2000 Conference on
Conference_Location :
Sydney, NSW, Australia
Print_ISBN :
0-7803-5744-2
Type :
conf
DOI :
10.1109/CPEM.2000.850875
Filename :
850875
Link To Document :