DocumentCode :
2131858
Title :
Traceable calibration of specific absorption rate (SAR) for mobile phone dosimetry
Author :
Liang, W. ; West, P. ; Loader, B. ; Lees, K. ; Gregory, A.P. ; Clarke, R.N.
Author_Institution :
Nat. Phys. Lab., Teddington, UK
fYear :
2000
fDate :
14-19 May 2000
Firstpage :
496
Lastpage :
497
Abstract :
This paper describes an NPL project for providing traceability for SAR measurement in mobile phone dosimetry. Electric field implantable probe calibration facilities have been developed for this purpose. Computer simulation has been used to assist in analysing the uncertainty budgets for these calibrations. A new international intercomparison has been proposed for measurement of this parameter.
Keywords :
UHF measurement; biological effects of fields; calibration; dosimetry; electric field measurement; probes; radiotelephony; 1800 MHz; 900 MHz; NPL project; SAR measurement; cell phones; electric field implantable probe; mobile phone dosimetry; probe calibration facilities; specific absorption rate; traceable calibration; Brain modeling; Calibration; Dielectric measurements; Dosimetry; Imaging phantoms; Mobile handsets; Permittivity measurement; Probes; Specific absorption rate; TEM cells;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2000 Conference on
Conference_Location :
Sydney, NSW, Australia
Print_ISBN :
0-7803-5744-2
Type :
conf
DOI :
10.1109/CPEM.2000.851100