DocumentCode
2133713
Title
Low power sensor calibration at KRISS
Author
Jeong-Hwan Kim ; Jeong-Il Park
Author_Institution
Div. of Electromagnetic Metrol., Korea Res. Inst. of Stand. & Sci., Taejon, South Korea
fYear
2000
fDate
14-19 May 2000
Firstpage
654
Lastpage
655
Abstract
For the power sensors to be used at below 1 mW power range has been implemented a new calibration method where thermistor mounts are used for the transfer of the power standard calibrated in a microcalorimeter.
Keywords
bolometers; calibration; calorimeters; electric sensing devices; measurement uncertainty; power measurement; thermistors; transfer standards; 1 mW; 1 mW power range; KRISS; measurement uncertainty; microcalorimeter; power sensors; power standard; thermistor mounts; transfer; Bolometers; Calibration; Directional couplers; Power generation; Power measurement; Reflection; Sensor phenomena and characterization; Thermal sensors; Thermistors; Voltmeters;