DocumentCode :
2138458
Title :
Highly Accurate Absolute Group Delay Measurement Technique for Frequency-Translation Devices
Author :
Le Truong, My ; Knox, Michael ; Dunsmore, Joel
Author_Institution :
Hewlett-Packard Company, Microwave Instruments Division, 1400 Fountaingrove Parkway, Santa Rosa, California 95403
Volume :
1
fYear :
1998
fDate :
Oct. 1998
Firstpage :
98
Lastpage :
103
Abstract :
This paper describes an advanced absolute group delay measurement technique. It uses a mixer calibration standard and vector error correction to achieve high-precision absolute group delay measurements of frequency-translation devices, such as mixers and frequency converters. The advantages of this technique include: 1. Improved absolute and relative group delay measurement accuracy of +/¿ 150 ps 2. Lower measurement ripples for highly accurate S-parameter measurements 3. Vector-calibrated performance by characterizing and removing test system errors This application is most valuable for designers, manufacturers, and system integrators of communication systems (such as telemetry applications, satellite, LMDS and point-to-point radio) in which high-accuracy group delay measurements are of great importance.
Keywords :
Calibration; Delay; Error correction; Frequency conversion; Frequency measurement; Measurement standards; Measurement techniques; Mixers; Scattering parameters; System testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Conference, 1998. 28th European
Conference_Location :
Amsterdam, Netherlands
Type :
conf
DOI :
10.1109/EUMA.1998.338099
Filename :
4139055
Link To Document :
بازگشت