• DocumentCode
    2157276
  • Title

    A numerical assesment of the effect of MRI surface coils on implanted pacemakers

  • Author

    Fontana, Nunzia ; Monorchio, Agostino ; Torrico, Max Ovidio Munoz ; Hao, Yang

  • Author_Institution
    Dipt. di Ing. dell´´Inf., Univ. of Pisa, Pisa, Italy
  • fYear
    2012
  • fDate
    8-14 July 2012
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    Electromagnetic interaction between a radio frequency surface coil for magnetic resonance imaging (MRI) and a heterogeneous human model implanted with a pacemaker system has been studied in order to provide safety requirements and to estimate potential damaging of the device. Since the coil can be considered an electrically small structure, an equivalent model has been considered and validated in order to reduce calculation burden in time domain simulation by using a Finite Integration Technique (FIT).
  • Keywords
    biological effects of fields; biomedical MRI; integration; pacemakers; FIT; MRI surface coil effect; device potential damaging estimation; electromagnetic interaction; finite integration technique; heterogeneous human model; implanted pacemaker system; magnetic resonance imaging; numerical assessment; radiofrequency surface coil; safety requirement; time-domain simulation; Coils; Computational modeling; Magnetic resonance imaging; Numerical models; Pacemakers; Radio frequency; Solid modeling;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Antennas and Propagation Society International Symposium (APSURSI), 2012 IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1522-3965
  • Print_ISBN
    978-1-4673-0461-0
  • Type

    conf

  • DOI
    10.1109/APS.2012.6349161
  • Filename
    6349161