DocumentCode :
2174662
Title :
Signal processing for a standard harmonic analyzer
Author :
Trigo, Leonardo ; Slomovitz, Daniel
Author_Institution :
UTE Laboratory, Montevideo, Uruguay
fYear :
2015
fDate :
24-27 Feb. 2015
Firstpage :
1
Lastpage :
3
Abstract :
This paper describes the signal processing used in the development of a power quality standard for calibrating commercial harmonic analyzers. High precision digitizers of 27 bits are used to get the requested accuracy, but they have low sampling rates. This drawback limits the maximum harmonic content that the analyzer can measure. This paper shows an application to avoid the Nyquist frequency limit allowing the computation of harmonic frequency up to twice that limit.
Keywords :
Accuracy; Bandwidth; Frequency measurement; Harmonic analysis; Power harmonic filters; Standards; Nyquist theorem; calibration; harmonic; high precision measurements; standard;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Circuits & Systems (LASCAS), 2015 IEEE 6th Latin American Symposium on
Conference_Location :
Montevideo, Uruguay
Type :
conf
DOI :
10.1109/LASCAS.2015.7250499
Filename :
7250499
Link To Document :
بازگشت