Title :
Multi-Shot Diffusion Weighted PROPELLER on 0.35T Open MRI System
Author :
Li, Jianqi ; Chen, Weibo ; Jiang, Xiaoping ; Wang, Yi
Author_Institution :
Shanghai Key Lab. of MR & Dept. of Phys., East China Normal Univ., Shanghai, China
Abstract :
The purpose of this project was to develop diffusion weighted periodically rotated overlapping parallel lines with enhanced reconstruction (DW-PROPELLER) on low-field open MR systems. The DW-PROPELLER was implemented on a 0.35T open MR system, and an auto pre-scan technique was used to mitigate non-Carr-Purcell-Meiboom-Gill (CPMG) artifacts. High quality diffusion weighted images and associated ADC maps were obtained. Multi-shot DW-PROPELLER is feasible for clinical use in the diagnosis of acute cerebral infarction on low-field open systems.
Keywords :
biodiffusion; biomedical MRI; brain; image reconstruction; medical image processing; neurophysiology; CPMG artifacts mitigation; acute cerebral infarction diagnosis; auto prescan technique; brain imaging; diffusion weighted periodically rotated overlapping parallel lines; enhanced image reconstruction; low-field open MRI system; magnetic flux density 0.35 T; multishot DW-PROPELLER; nonCarr-Purcell-Meiboom-Gill artifacts; Blades; Brain; Diffusion tensor imaging; Image reconstruction; Magnetic resonance imaging; Open systems; Physics; Propellers; Radiology; Sampling methods;
Conference_Titel :
Biomedical Engineering and Informatics, 2009. BMEI '09. 2nd International Conference on
Conference_Location :
Tianjin
Print_ISBN :
978-1-4244-4132-7
Electronic_ISBN :
978-1-4244-4134-1
DOI :
10.1109/BMEI.2009.5305028