  • Title

    Automated calibration methods for a high-voltage Park divider

  • Author

    Tsung-Han Su ; I-Pao Chen ; Tsao, S.H.

  • Author_Institution
    Nat. Meas. Lab., Hsinchu, Taiwan
  • fYear
    1994
  • fDate
    June 27 1994-July 1 1994
  • Firstpage
    307
  • Lastpage
    308
  • Abstract
    We describe two new methods of calibrating a 100-kV Park divider, fitted with N/10 taps. They prescribed a series of automated measurements using (i) a digital conductance comparator, and (ii) a set of full-bridge circuits configurable entirely from the divider sections. Self-heating effects were studied systematically under PC control. Preliminary measurements determined the 1/10 ratio to /spl plusmn/25 ppm.<>
  • Keywords
    bridge circuits; calibration; comparators (circuits); computerised instrumentation; electric sensing devices; high-voltage techniques; voltage dividers; voltage measurement; 100 kV; PC control; automated calibration; digital conductance comparator; full-bridge circuits; high-voltage Park divider; self-heating effects; Automatic control; Bridge circuits; Calibration; Control systems; Digital relays; Integrated circuit measurements; Laboratories; Microcomputers; Testing; Voltage control;