DocumentCode :
2187501
Title :
Comparison between a vector multiport network analyzer and the national S-parameter: measurement system
Author :
Ferrero, A. ; Sanpietro, F. ; Pisani, U. ; Brunetti, L.
Author_Institution :
Politecnico di Torino, Italy
fYear :
1994
fDate :
June 27 1994-July 1 1994
Firstpage :
143
Lastpage :
144
Abstract :
A multiport vector network analyzer based on a new calibration concept, has been compared with the P-port S-parameter National measurement system at IENGF. The measurements were performed on precision 7 mm standard components and exhibited an optimum agreement. These results open the possibility to use the new multiport network analyzer for certification measurements of multiport devices.<>
Keywords :
S-parameters; calibration; measurement standards; multiport networks; network analysers; IENGF; P-port S-parameter National measurement system; certification measurement; national S-parameter; vector multiport network analyzer; vector network analyzer; Attenuators; Calibration; Connectors; Frequency measurement; Impedance; Logic testing; Measurement standards; Performance evaluation; Scattering parameters; System testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements, 1994. Digest., 1994 Conference on
Conference_Location :
Boulder, CO, USA
Print_ISBN :
0-7803-1984-2
Type :
conf
DOI :
10.1109/CPEM.1994.333411
Filename :
333411
Link To Document :
بازگشت