DocumentCode :
2189920
Title :
Research on tipping calibration and exponent calibration based on LN2 for ground-based multi-channel microwave radiometer
Author :
He Jieying ; Zhang Shengwei ; Zhang Yu
fYear :
2012
fDate :
22-27 July 2012
Firstpage :
4652
Lastpage :
4655
Abstract :
Respect to ground-based microwave radiometer, the paper presents exponent calibration based on blackbody cooled by liquid nitrogen. In order to correct the calibration coefficients, this paper also presents tipping method at low optical depth (K band), which is different from common tipping method. It observes the clear-sky and clear-sky with noise injection to correct nonlinear factor and stability of noise. The calibration results show that using tipping method can achieve high resolution.
Keywords :
calibration; microwave measurement; radiometers; stability; LN2; exponent calibration; ground-based multichannel microwave radiometer; liquid nitrogen; low optical depth; nonlinear factor; tipping calibration; tipping method; Atmospheric measurements; Calibration; Microwave measurements; Microwave radiometry; Microwave theory and techniques; Noise; Temperature measurement; exponent calibration; microwave radiometer; noise injection; tipping calibration;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Geoscience and Remote Sensing Symposium (IGARSS), 2012 IEEE International
Conference_Location :
Munich
ISSN :
2153-6996
Print_ISBN :
978-1-4673-1160-1
Electronic_ISBN :
2153-6996
Type :
conf
DOI :
10.1109/IGARSS.2012.6350428
Filename :
6350428
Link To Document :
بازگشت