DocumentCode :
2190040
Title :
Germanium strip detector Compton telescope using three dimensional readout
Author :
Wulf, Eric A. ; Johnson, W. Neil ; Kroeger, Richard A. ; Kurfess, James D. ; Phlips, Bernard F.
Volume :
1
fYear :
2002
fDate :
10-16 Nov. 2002
Firstpage :
57
Abstract :
Compton telescopes using two germanium strip detectors with depth resolution have been demonstrated at the Naval Research Lab. Depth resolution allows interactions to be located to less than 1 mm, down from 1 cm with no depth resolution, which improves the imaging resolution of the telescope substantially. Compton images and reconstructed energy spectra of events in which gamma rays interact three times in the two detectors but did not deposit their full energy (i.e. Three-Compton) are examined. Finally, the multiple modes of the system including a standard Compton telescope and a Three-Compton telescope are compared.
Keywords :
gamma-ray detection; germanium radiation detectors; image reconstruction; nuclear electronics; position sensitive particle detectors; readout electronics; 1 mm; Ge; depth resolution; gamma rays; germanium strip detector Compton telescope; imaging resolution; multiple modes; reconstructed energy spectra; three dimensional readout; Detectors; Energy resolution; Event detection; Gamma rays; Germanium; Image reconstruction; Image resolution; Optical imaging; Strips; Telescopes;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2002 IEEE
Print_ISBN :
0-7803-7636-6
Type :
conf
DOI :
10.1109/NSSMIC.2002.1239268
Filename :
1239268
Link To Document :
بازگشت