DocumentCode :
2193685
Title :
Multichannel MCG Systems with Optimum Combinations of Pickup Coils and Shielded Rooms
Author :
Lee, Y.H. ; Yu, K.K. ; Kim, K. ; Kim, J.M. ; Kang, C.S. ; Kwon, H. ; Lim, H.K. ; Park, Y.K.
Author_Institution :
Korea Res. Inst. of Stand. & Sci, Daejeon
fYear :
2007
fDate :
12-14 Oct. 2007
Firstpage :
297
Lastpage :
300
Abstract :
In order to provide optimum signal-to-noise ratios depending on the shielding conditions, we fabricated several multichannel magnetocardiogram systems with different types of pickup coils, magnetometer, first-order gradiometers and second-order gradiometers. The magnetocardiogram (MCG) systems were operated inside different magnetically shielded rooms, and signal-to-noise ratios were compared. The MCG systems have numbers of sensors in the range of 61-64, whose coverage diameters are large enough to measure the MCG maps in a single position setting. Considering the high cost of Permalloy needed in the shielded room, combination of a second-order gradiometer system and a thin magnetically shielded room seems best choice for the detection of ischemia in noisy hospital environment, like emergency department.
Keywords :
biomedical equipment; diseases; magnetic shielding; magnetocardiography; magnetometers; gradiometers; ischemia detection; magnetocardiogram; magnetometer; multichannel MCG systems; pickup coils; shielded rooms; Coils; Costs; Ischemic pain; Magnetic sensors; Magnetic shielding; Magnetometers; Position measurement; Sensor systems; Signal to noise ratio; Working environment noise;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Noninvasive Functional Source Imaging of the Brain and Heart and the International Conference on Functional Biomedical Imaging, 2007. NFSI-ICFBI 2007. Joint Meeting of the 6th International Symposium on
Conference_Location :
Hangzhou
Print_ISBN :
978-1-4244-0949-5
Electronic_ISBN :
978-1-4244-0949-5
Type :
conf
DOI :
10.1109/NFSI-ICFBI.2007.4387758
Filename :
4387758
Link To Document :